Pular para o conteúdo principal
GE Healthcare · MRExcite2223621

Internal Software Errors due to Network Switch.

Symptom

Unable to Scan with Internal Software Errors 2223621 lx-mr scn mrmail_connect_mgr.c 239 Exception Class: Unknown Function: Software/Firmware: Host Communications Internal Software Error. Can't callback on undefined NMRID:COILCK.

Error message

2223621 Internal Software Error. Can't callback on undefined NMRID:COILCK. 2223621 lx-mr scn mrmail_connect_mgr.c 239 Exception Class: Unknown Function: Software/Firmware: Host Communications Internal Software Error. Can't callback on undefined NMRID:COILCK.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT.
Area
MR

Full cause & step-by-step resolution