CWA ID121877. TPS not responding during applications boot with Error 2223817.

Symptom

TPS not responding during applications boot with Error 2223817 See long description and attached file. NO errors in mgd_term windows. All three processors finish boot with applications startup complete message. /usr/g/service/log/scn.out file showed MGD chassis failed boot. See attached file.

Error message

2223621 Internal Software Error. Can''t callback on undefined NMRID:ps_exec. During TPS Reset: 1081934445 0 1 Wed Apr 14 09:20:45 2004 2223621 4 CARD psc mrmail_connect_mgr.c 238 Internal Software Error. Can''t callback on undefined NMRID:ps_exec.

2227996 AP initialization error Auto configuration failed. The number of CE nodes detected is 0 During TPS Reset: Wed Apr 14 09:20:40 2004 2227996 CARD NSP APS:ap_init 199 AP initialization error: Auto configuration failed. The number of CE nodes detected is 0.

2223625 Send packet opcode:814 seqnum:9 rev:2 type:1 length:328 req:psc_major resp:ps_exec status:0 failed. During TPS Reset: 1081934445 1 1 Wed Apr 14 09:20:45 2004 2223625 4 CARD psc mrmailmgr.c 791 Send packet opcode:814 seqnum:9 rev:2 type:1 length:328 req:psc_major resp:ps_exec status:0 failed.

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

Applies to

Models
Excite, Ovation
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.