The Customer complains of a Water or a fluid leak on the system.

Symptom

The Customer complains of a Water or a fluid leak on the system. The customer indicates there is water leaking from the closed loop cooling system. Fluid is coming out of the tube arm assembly or in the technical room. If zero stock on 2184070-2 RP ROTATING JOINT then order 2129844-2 RIGID WATER COOLING ASSEMBLY and it contain the joint part. Also replace the two O-rings with new or they will leak next. The Customer complains of a Water or a fluid leak on the system. The customer indicates there is water leaking from the closed loop cooling system. Fluid is coming out of the tube arm assembly or in the technical room. 2184070-2 RP ROTATING JOINT then order 2129844-2 RIGID WATER COOLING ASSEMBLY and it contain the joint part. Also replace the two O-rings with new or they will leak next

Applies to

Models
Innova, Advantx
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems
Component
Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution