Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova10706721070673

IQ+ or PRO table, "!!

Symptom

IQ+ or PRO table, "!! Patient fall risk. Tabletop brakes failure. Handle patient carefully (2.9)" IQ+ or PRO table, system displays an error message on the monitor that reads: "!! Patient fall risk. Tabletop brakes failure. Handle patient carefully (2.9)" IQ+ or PRO table, Elegance system displays an error message on the monitor that reads: "!! Patient fall risk. Tabletop brakes failure. Handle patient carefully (2.9)" Started with IQ+ or PRO table, ELEGANCE_IS2, DL_ELEGANCE_3.2.1

Error message

INX100: Elegance 1070672 or 1070673 (persistent weak brake restored at boot) INX100: Elegance 1070672 (persistent weak brake restored at boot). See long description for complete error pattern. May have a similar pattern starting by 1070673 error (related to lateral weak brake failure) Persistent lat weak brake error restored at boot" WR Proactive Digital Service did report following pattern: Note : May have a similar pattern starting by 1070673 error (related to lateral weak brake failure) Persistent lat weak brake error restored at boot" WR

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
5269338-3. Servo node FW version 3.78 Servo node FW from version 3.78. The kit part number is 5269338-3 (Elegance sw package which includes correct servo node release 3.78). Servo node FW from version 3.78. The kit part number is 5269338-3 (Elegance sw package which includes correct servo node release 3.78). Always use the most appropriate documentation for your system. The following FRU reference was taken from Direction 5192425-2-2EN Rev#1 and is provided for your reference from the Innova 2100-IQ, 3100-IQ, 4100-IQ Innova-IQ DVD documentation. To view the most recent appropriate and detailed documentation, go to MY WORKSHOP
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues