GE Healthcare · X-ray VascularInnova10706721070673
IQ+ or PRO table, "!!
Symptom
IQ+ or PRO table, "!! Patient fall risk. Tabletop brakes failure. Handle patient carefully (2.9)" IQ+ or PRO table, system displays an error message on the monitor that reads: "!! Patient fall risk. Tabletop brakes failure. Handle patient carefully (2.9)" IQ+ or PRO table, Elegance system displays an error message on the monitor that reads: "!! Patient fall risk. Tabletop brakes failure. Handle patient carefully (2.9)" Started with IQ+ or PRO table, ELEGANCE_IS2, DL_ELEGANCE_3.2.1
Error message
INX100: Elegance 1070672 or 1070673 (persistent weak brake restored at boot) INX100: Elegance 1070672 (persistent weak brake restored at boot). See long description for complete error pattern. May have a similar pattern starting by 1070673 error (related to lateral weak brake failure) Persistent lat weak brake error restored at boot" WR Proactive Digital Service did report following pattern: Note : May have a similar pattern starting by 1070673 error (related to lateral weak brake failure) Persistent lat weak brake error restored at boot" WR
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- 5269338-3. Servo node FW version 3.78 Servo node FW from version 3.78. The kit part number is 5269338-3 (Elegance sw package which includes correct servo node release 3.78). Servo node FW from version 3.78. The kit part number is 5269338-3 (Elegance sw package which includes correct servo node release 3.78). Always use the most appropriate documentation for your system. The following FRU reference was taken from Direction 5192425-2-2EN Rev#1 and is provided for your reference from the Innova 2100-IQ, 3100-IQ, 4100-IQ Innova-IQ DVD documentation. To view the most recent appropriate and detailed documentation, go to MY WORKSHOP
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.