The Innova console displays Digital or keypad failure, reset system and call Service
Symptom
IN2K: The Innova console displays Digital or keypad failure, reset system and call Service The Innova console displays Digital or keypad failure, reset system and call Service. 2800015 "Due to a possible communication error detected with PDU, power supply is not controlled" 1. Run DL subsystem/ Interactive/PDU diagnostics. 2. If no error found then replace ATLAS PDU 4KVA WITH AC REG- WITH UPS 2800013 Due to a possible hardware error detected with PDU, power supply is not controlled" 1. Run DL subsystem/ Interactive/PDU diagnostics. 2. If no error found then replace ATLAS The Innova console displays Digital or keypad failure, reset system and call Service May see in error log for Innova 2000 2800015 "Due to a possible communication error detected with PDU, power supply is not controlled" Run DL subsystem/ Interactive/PDU diagnostics. If no error found then replace ATLAS PDU 4KVA WITH AC REG- WITH UPS 2800013 Due to a possible hardware error detected with PDU, power supply is not controlled" Run DL subsystem/ Interactive/PDU diagnostics. If no error found then replace ATLAS PDU 4KVA WITH AC REG- WITH UPS 2800016 "Due to a possible hardware error detected with the local Keypad, local Keypad is not available" Run DL subsystem/Interactive/ Zephir serial diagnostics with loopback succesively on DL and on cable. If no error found, then replace DLX KEYPAD
Error message
2800020: Root Cause: Component(s) - DL Display board This error is due to asynchronous error found on the Display board.
Applies to
- Models
- Innova, Advantx
- Product
- X Innova 4100 X Do not use---------4100 systems 4100 systems
- Software
- Bolus M4+ DL release : DL_LBYBOLUSM4_10.1.050805 Ghost DL-2 : NA Ghost DL-3 : 18.8 Ghost DL-5 : 20.6 DL : DL_LBYBOLUSM4_10.1.050805 DL Atlas Core : 3.1.R RTAC Atlas Core : 3.1.R IDC : 4.5.5 Calypso : 2.21 Calypso Uptime : BOLUS1.24 Positioner : 1.2.24B_IS6.6 Jedi : B6.30BFC
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.