GE Healthcare · X-ray VascularInnova12001021200107130010113001071300112

Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Image Monitor.

Symptom

Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Image Monitor. Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Monitor.

Error message

SMC detector Conditioner ERR11 is displayed on the LCD SMC detector Conditioner ERR11 is displayed on the LCD. ERR 11 indicates a dc power malfunction has been deteected by the conditioner. Power cycle the chiller, if no error state returns, chiller is ok. Check for external chiller damange and any excessive noise. If error state returns, Replace Chiller.

INX100: 1200102 Failed to send boot parameters to the conditioner followed by 1200107 conditioner boot failure. 1200102 ERR CONDITIONER DO XRImDet DEM Component(s) Conditioner 1 0 [DemSupervisor:: bootConditioner] Failed to send boot parameters DemSupervisor.cpp 385 1200107 ERR CONDITIONER BOOT FAILURE IH XRImDet DEM Component(s) Conditioner 1 0

1200107 (Frontal) 1300101 (Lateral) ERR_Conditioner_Boot_Failure 1200107 (Frontal) 1300101 (Lateral) ERR_Conditioner_Boot_Failure. 1200107 (1300101): Root Cause: Component(s) - Frontal (Lateral) Conditioner This error is due to time out on Communication to the frontal (lateral) Detector Conditioner or bad status returned by Conditioner. 1200107 (Frontal) - 1300101 (Lateral) ERR_Conditioner_Boot_Failure 1200107 (1300101) : Root Cause: Component(s) - Frontal (Lateral) Conditioner This erro

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Detector conditioner 5131740 SMC detector conditioner Detector conditioner Current rev is -15 but a -7 through -15 is acceptable. Detector Conditioner, connecter kit 2379172 5131740 SMC detector conditioner Detector conditioner Current rev is -15 but a -7 through -15 is acceptable. Detector Conditioner, connecter kit 2379172
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.