GE Healthcare · X-ray VascularInnova400005440001014000048
Innova 2000 did not fully boot.
Symptom
Innova 2000 did not fully boot. DL is stuck on mire pattern, advantx console is blank. innova 2000 did not fully boot. DL is stuck on mire pattern, advantx console is blank. Innova 2000 did not fully boot. DL is stuck on mire pattern, advantx console is blank.
Error message
In2K, several communication errors inthe log including:4020- 7, 4000054, 4000101, 4000048 7 IH CONSOLE SYSTEM ERROR: NOT COMMUNICATING WITH NODE Debug: fct-status=3 hw-status=0 pos-error=4020 usr-status=0 HWStatus=0000 User=0003 0000 4020 4000054 AR XRTitan connect call failed Debug: null 4000101 AR XRTitan Select call failed Debug: null 4000048 AR XRTitan could not send packet Debug: lost connection and could not reconnect 4020-7 IH CONSOLE SYSTEM ERROR: NOT COMMUNICATING WITH NODE Debug: fct-status=3 hw-status=0 pos-error=4020 usr-status=0 HWStatus=0000 User=0003 0000 4020 4000054 AR XRTitan connect call failed Debug: null 4000101 AR XRTitan Select call failed Debug: null 4000048 AR XRTitan could not send packet Debug: lost connection and could not reconnect
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems
- Component
- advantx console power supply. 210938055W 4-OUTPUT SWITCHING POWER SUPPLY advantx console power supply. 210938055W 4-OUTPUT SWITCHING POWER SUPPLY advantx console power supply. 210938055W 4-OUTPUT SWITCHING POWER SUPPLY
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.