Innova 2000 did not fully boot.

Symptom

Innova 2000 did not fully boot. DL is stuck on mire pattern, advantx console is blank. innova 2000 did not fully boot. DL is stuck on mire pattern, advantx console is blank. Innova 2000 did not fully boot. DL is stuck on mire pattern, advantx console is blank.

Error message

In2K, several communication errors inthe log including:4020- 7, 4000054, 4000101, 4000048 7 IH CONSOLE SYSTEM ERROR: NOT COMMUNICATING WITH NODE Debug: fct-status=3 hw-status=0 pos-error=4020 usr-status=0 HWStatus=0000 User=0003 0000 4020 4000054 AR XRTitan connect call failed Debug: null 4000101 AR XRTitan Select call failed Debug: null 4000048 AR XRTitan could not send packet Debug: lost connection and could not reconnect 4020-7 IH CONSOLE SYSTEM ERROR: NOT COMMUNICATING WITH NODE Debug: fct-status=3 hw-status=0 pos-error=4020 usr-status=0 HWStatus=0000 User=0003 0000 4020 4000054 AR XRTitan connect call failed Debug: null 4000101 AR XRTitan Select call failed Debug: null 4000048 AR XRTitan could not send packet Debug: lost connection and could not reconnect

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Component
advantx console power supply. 210938055W 4-OUTPUT SWITCHING POWER SUPPLY advantx console power supply. 210938055W 4-OUTPUT SWITCHING POWER SUPPLY advantx console power supply. 210938055W 4-OUTPUT SWITCHING POWER SUPPLY
Area
X-ray Vascular