Pular para o conteúdo principal
GE Healthcare · MRExcite224485822445932244594224485522445852244586

SGA Power Supply Internal Wiring Fault or Control Undervoltage Due to SGA Power Supply.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244858 The GP is reporting SGA Power Supply Internal Wiring Fault or Control Undervoltage. the same error as 2244593 on 11X, G3 and 2244594 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5. 2244858 error code on Post 14M5 is the same error as 2244593 on 11X, G3 and 2244594 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5. 2244858 The GP is reporting SGA Power Supply Internal Wiring Fault or Control Undervoltage.

2244855 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready. Same as 2244585 and 2244586. 2244855 errorcode on Post 14M5 is the same error as 2244585 on 11X, G3 and 2244586 on 12X, E2, 14M4 (HDx, HDe). 2244855 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready. 2244855 errorcode on Post 14M5 is the same error as 2244585 on 11X, G3 and 2244586 on 12X, E2, 14M4 (HDx, HDe).2244855 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues