GE Healthcare · X-ray VascularInnovaAdvantx1400005
System can't boot up with error 4F2E Reports on Advantx Console;:Innova 2000:1400005
Symptom
System can't boot up with error 4F2E Reports on Advantx Console; message "Xray not available; Reset system to recover; Xray; Call Service" display on console
Error message
1400005 ERR DATA BASE IH XRImPreproc uponFPGAInitialized File Corruption Application 1 0 [Preprocessing::readImageOrientationConfig] could not find /c/prod/config/imprep/orient.txt preprocessing.cpp 741 1400005 ERR DATA BASE IH XRImPreproc uponFPGAInitialized File Corruption Application 1 0 [Preprocessing::readImageOrientationConfig] could not find /c/prod/config/imprep/orient.txt preprocessing.cpp 741
Applies to
- Models
- Innova, Advantx
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems
- Software
- $, Innova 2000 H2VCP M4 FMI, DL release M4_12.17_REV1 M4_12.17_REV1 DL release 12.32.2 Ghost DL-1 "F" : NA Ghost DL-1 "G" : NA Ghost DL-2 : NA Ghost DL-4 : 19.9 DL : 12.32.2 DL Atlas Core : 3.1Q RTAC Atlas Core : 3.1O-1 IDC : 1.15 Calypso : 13.6 Titan : H0.58 Pos. Motorola : NA Pos. Roadrunner : 5.2K-2 Calypso Uptime : 12.17_Rev1
- Component
- 5129430ATLAS PC 80GB SYSTEM DISK ASSY RP 5129430ATLAS PC 80GB SYSTEM DISK ASSY RP 5129430ATLAS PC 80GB SYSTEM DISK ASSY RP
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.