Air Conditioning failure in the Computer room.

Symptom

The equipment room is very hot. Customer has called for AC repair, the computer room temp is 93 F. The control room and scan room temps appear to be good.

Error message

2248609 The air temperature in the System Cabinet is greater than 50'C (122F) Mon Aug 22 12:32:49 2005 Error: 2248609 Host: MR3T Process: NSP File: SCP:IrfPciHWO Line: 1234 The air temperature in the System Cabinet is greater than 50'C (122F). Check equipment room temp, if warmer than normal contact site maintenance, else call GE Service. diags error is 2247127

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.