Pular para o conteúdo principal
GE Healthcare · MRSignaExcite22473836851142247340

error 2247383 The signal's channel-to-channel variance exceeded the specification for channel Due to UTNS3

Symptom

16 Channel Switch Diagnostics Fail

Error message

2247383 The signal's channel-to-channel variance exceeded the specification for channel: Test 16 Channel Switch Output Osc See Long Description Error: 2247383 Host: b5 Process: diagCoord File: rrfRxFruIso.c Line: 2796 Diagnostic: 16 Channel Switch Iteration: 1 of 1 Action: 1 Task ID: 0x7f53e50 Time: 685114 Object: APS Test Task Function: Software/Firmware/Config: N/A Test: '16 Ch Switch Output Osc. OB(O)' diagnostic error. The signal's channel-to-channel variance exceeded the specification for channel: 7, 8.-> Extended Error: Review the data in /usr/g/service/datamine. The summary file contains the results used for logging this error. If the file exists, the raw and processed data is available in a corresponding data file whose name contains the test iteration tag.

2247340 Exciter calibration diagnostics failed. Please perform a TPS reset.

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2366218-2 UTNS3 See Long Description 2366218-2 UTNS3 EMC ED2 Compliant .
Area
MR