Scan stops with gradient error 2244897 The GP is reporting a % Axis OverTemperature Fault due to the HFA Control Board.

Symptom

Scan stops with gradient error 2244897.

Error message

2244897 The GP is reporting a % Axis OverTemperature Fault See Long Description System will wait 5 minutes before attempting recovery. Extended Error: Check that the cabinet fan is functioning properly, rear door is closed and sealed, and that no filler panels have been removed from the front of the cabinet. Proper cooling requires a vaccum to pull air through the Gradient Amplifiers, Power Supply, and GP. If fault persists on specific PSDs write a PQR. If fault persists on many or all PSDs, replace Gradient Amplifier.

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2396050-2 HFA CONTROL BOARD ASSEMBLY
Area
MR

Full cause & step-by-step resolution