Body and Head Maximum Power RF Output Check Calls For Part Number 5166139, 7/16(M)-N(F) Adapter, But It Isn't Orderable.…

Body and Head Maximum Power RF Output Check Calls For Part Number 5166139, 7/16(M)-N(F) Adapter, But It Isn't Orderable. 7/16(M)-N(F) Adapter, Part Number 5166139 is Included With the System Cabinet i…

Symptom

Body and Head Maximum Power RF Output Check Calls For Part Number 5166139, 7/16(M)-N(F) Adapter, But It Isn't Orderable See Long Description The Body and Head Maximum Power RF Output Check procedure in the Excite HDe Service Methods in section 3.1 "Tools and Test Equipment', calls for part number 5166139, 7/16(M)-N(F) Adapter, but it isn't in the RF Power Measurement Kit, it isn't a FRU, and it's not orderable. The procedure, in step 4.4.2.a states as follows "An RF adapter is provided in the RF Power Measurement Kit to connect between the 7/16 J4 body RF output and the RF Power Measurement Kit 40 dB N-connector coupler." But this isn't true because the FE could not find that part in the kit. How does the FE get this part so he can complete the calibration? This procedure is found in Direction 5323683-2EN, Rev# 4.0 Signa HDe 1.5T Service Methods Module Version 5.0, Version Date 7/21/09 Under Adjust/Cals --> RF System --> 1.5T SRFD3 Max Power RF Out Setup and Calibration To view the most recent appropriate and detailed documentation, click to go to MR Service Documentation.

Applies to

Models
Excite, Signa
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.