Emergency Error Log Viewer or CSD VNC (remote SUIF) comes up with blank screen and/or a tcp error

Symptom

ISD user only, How to reset Emergency Error Log Viewer or CSD VNC (remote SUIF) if it comes up with blank screen and/or a tcp error ISD user only, How to reset Emergency Error Log Viewer or CSD VNC (remote SUIF) if it comes up with blank screen and/or a tcp error The Iipadmin web server hung. Run the command iipadmin stop start, wait for the prompt to return, and try again insite> iipadmin stop start The IIPWebServer service is stopping... The IIPWebServer service was stopped successfully. The IIPWebServer service is starting. The IIPWebServer service was started successfully insite>

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Software
$ Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9 FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 healthpage tag for this software is UNITY_M4_FMI_4.21 212009270 03/26/2010 13:31 GMT
Area
X-ray Vascular