Pular para o conteúdo principal
GE Healthcare · MRSignaOvationExcite2249107

CWA SCP/SRI Link has Failed Attempt to Recover Error on Scan page.

Symptom

SCP/SRI Link has Failed Attempt to Recover Error on Scan page

Error message

2249107 SRI SCP has recieved a packet indicating the SRI has reset SRI heart beat is steady on 2249107 SRI SCP has recieved a packet indicating the SRI has reset SRI heart beat is steady on

Applies to

Models
Signa, Ovation, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues