GE Healthcare · X-ray VascularInnova3602047

Xray unavailable message on live display with fluoro and record modes with footswitch depressed.

Symptom

Xray unavailable message on live display with fluoro and record modes with footswitch depressed. System must be shutdown and rebooted to recover as a reset will not recover. May be able to produce radiation Fluoro and Record, X-ray Tone and indicator are on but "X-ray Unavailable" or "X-ray Abort" Message on Screen and no Image on Monitor. This could also cause a Database Inconsistency error because the RTAC is sending the acquired image data to the DL but the PPC bd in the DL is not receivin Xray unavailable message on live display with fluoro and record modes with footswitch depressed. System must be shutdown and rebooted to recover as a reset will not recover. May be able to produce radiation Fluoro and Record, X-ray Tone and indicator are on but "X-ray Unavailable" or "X-ray Abort" Message on Screen and no Image on Monitor. This could also cause a Database Inconsistency error because the RTAC is sending the acquired image data to the DL but the PPC bd in the DL is not receiving because the FCIB in DL has locked up or failed. Hence no image on Monitor and Data was not written to disk which could generate a Database Inconsistency.

Error message

error Code = 3602047, error Text = DC bus out of range, debug Text += jedi-error=501h, DC bus out of range, recovery Class = IH, major Function /= XRTechMgr System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service".

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Software
$ M4 SPAD BIP_4.18.5, DL_SPAD_4.15.1 FMI 12108 $New, M4 SPAD BIP_4.18.5, DL_SPAD_4.15.1 FMI 12108 or $ new, Innova Biplane M4 BIP_4.18.5, DL_4.15.1 FMI 12108 $New, M4 SPAD BIP_4.18.5, DL_SPAD_4.15.1 FMI 12108 or $ new, Innova Biplane M4 BIP_4.18.5, DL_4.15.1 FMI 12108
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.