GE Healthcare · MRExcite22547932254753150811300022547862254781
Unable to Download or Reset The TPS MGD or ISE.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2254793 The Universal Power Monitor 2 failed its DAC Self Test on the Narrow Band Head Channel Error: 2254793 Host: ECH-GEMR Process: NSP File: SCP:UPM Line: 1059 The Universal Power Monitor 2 failed its DAC Self Test on the Narrow Band Head Channel. Expected Data : 0x2c00. Detected Data : 0x4000.-> 2254753 The Universal Power Monitor 2 Has Detected a Power Supply Fault Error: 2254753 Host: ECH-GEMR Process: NSP File: SCP:UPM Line: 1553 The Universal Power Monitor 2 Has Detected a Power Supply Fault.^A Source of Fault is UPB -12 Volts. Actual Voltage = 15081 mV Limit Voltage = 13000 mV-> 2254786 The Universal Power Monitor 2 Has Detected an Invalid Configuration. Source of Fault is NB Detector Board RF Cable Missing. See Long Description for full error and extended log... NOTE: The errors below were copied from a 3.0T 14xM5 system: Date Time ErrorCode Process File Line Message Jan 24 Thu 17:11:21 2254786 NSP SCP:UPM 894 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: Unknown Coil: Unknown Scan: -1/-1 The Universal Power Monitor 1 Has Detected an Invalid Configuration. Source of Fault is NB Detector Board RF Cable Missing.-> Jan 24 Thu 17:11:21 22547
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Component
- 5250034 RF Detector 2 Board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.