Elegance table 1070573 Lift does not work on Boot.

Symptom

Elegance table 1070573 Lift does not work on Boot. The table is operating in the degraded mode. When the system is booted in the morning the table fails the lift PRDs . The table is operating in the degraded mode. Swapped the Lift Servo Node with the Tilt. The error follows the Servo Node. There is no need to reload SW in that all Servo Node modules use the same SW. The calibration data is loaded from the table CPU on boot.

Error message

Elegance table 1070573 PRD Error: ADC test failed for lift axis Lift does not work on Boot. The table is operating in the degraded mode. Error codes 1050297, 1070573, 1070384, 1070581,1070161 1070383 Elegance table 1070573 PRD Error: ADC test failed for lift axis The Elegance table Lift does not work on Boot. The table is operating in the degraded mode. Error codes 1070573, then other errors 1070384, 1070581,1070161 1070383. . The table is operating in the degraded mode.

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Area
X-ray Vascular

Full cause & step-by-step resolution