Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova1800013200013

Bad Pixel Calibration fails during PM Calibration.

Symptom

Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76

Error message

1800013 ERR C80 COM stopPrepare [DigAbdDriver::waitForC80Reply] timed out. May see 200013 after the time out 1800013 ERR C80 COM AR XRExpSens stopPrepare Component(s) RTAC C80 Boards 1 0 [DigAbdDriver::waitForC80Reply] timed out !!!! dig abd driver.cpp 68 May see 200013 after the time out

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Software
$, Innova 2000 Service Pack FMI F4, HerSvcPack_2.02 HerSvcPack_2.02 DL Applications version 11.23.26 Ghost DL-1 "F" : 17.4 Ghost DL-1 "G" : 15.5 Ghost DL-2 : 16.4 Ghost DL-4 : NA DL : 11.23.26 DL Atlas Core : 3.1O-1 RTAC Atlas Core : 3.1O-2 IDC : 2.13 Calypso : 11.13_FMI_7 Calypso Uptime : 11.25_FMI Titan : H0.57 Pos. Motorola : 5.2Y3-2 Pos. Roadrunner : 5.0R
Component
DL2 Chassis Cables in front, Single Ethernet Board, Video and Ethernet on CPU, Serial port 1 & 2 on bulkhead Front View Rear View Component Locations
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues