GE Healthcare · MR2239991
One of the Body Coil Bias boards reads open across the Pin Diode.
Symptom
Unable to Scan with Coil Bias Error
Error message
2239991 Coil Switch Bias Board error Error 2239991 Host: oc001 Proc: NSP File: HPC_PORT_ID:spi_srp Coil switched bias board error. Body coil disable error 1 occurred. Check the CSB Board, Coil ID or the diode switch of the body coil was broken. Error 2239991 Host: oc001 Proc: NSPFile: HPC_PORT_ID:spi_srpCoil switched bias board error.Body coil disable error 1 <OPEN> occurred. Check the CSB Board, Coil ID or the diode switch of the body coil was broken.
Applies to
- Product
- Contour 3 & 4
- Software
- Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
- Component
- 211552 and 212447 Bias Disable Assy's 211552 is not the same as 212447. 211552 is I channel bias disable board and 212447 is Q channel bias disable board. Order one of each, must be ordered as a pair.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.