GE Healthcare · MRExciteSigna225405750206
Need 12 in cable stub to complete Body Reflected Power Calibration (For 3.0T systems with LCC300 magnets).
Symptom
3T Rf amp UPM cal procedure section 4.3 body reflected power calibration calls for 12 in. cable stub, its not in the RF power measurement kit.
Error message
2254057 RF Lock signal is asserted This may be the result of a power monitor (UPM) fault or the RF amplifier may need to be reset. See Long Description 2254057 Host: WBH_RAD_MR11 Process: NSP File: SCP:Srfd3RfNbCanHWO Line: 2612 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Spin Echo Coil: BODY Scan: 50206/1 RF Lock signal is asserted. This may be the result of a power monitor (UPM) fault or the RF amplifier may need to be reset.
Applies to
- Models
- Excite, Signa
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 5250034 RF Detector 2 Board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.