Pular para o conteúdo principal
GE Healthcare · MRExciteSigna10102

Cant Restore Images From MOD Due To MOD Drive.

Symptom

Can Archive, but not Restore Images from MOD See Long Description FE's system has no problem archiving to media, but it cannot read any media after archiving them...always tells him disks are corrupt. He can take the same disk and read it on another system...no problem.

Error message

10102 Archive detects corrupt media. Wed Apr 20 07:49:10 2005 Error: 10102 Host: MRS1 Process: arsjob rchiveDevice.c Line: 4780 Archive detects corrupt media.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2318598 Sony MOD Drive Model SMO-F551-SD Sony MOD Drive 2318598 (Model #SMO-F551-SD) Sony MOD Drive 2318598 (Model #SMO-F551-SD)
Area
MR