GE Healthcare · X-ray VascularInnova120010713001011200112
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Symptom
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Error message
1200107 (Frontal) 1300101 (Lateral) ERR_Conditioner_Boot_Failure 1200107 (Frontal) 1300101 (Lateral) ERR_Conditioner_Boot_Failure. 1200107 (1300101): Root Cause: Component(s) - Frontal (Lateral) Conditioner This error is due to time out on Communication to the frontal (lateral) Detector Conditioner or bad status returned by Conditioner. 1200107 (Frontal) - 1300101 (Lateral) ERR_Conditioner_Boot_Failure 1200107 (1300101) : Root Cause: Component(s) - Frontal (Lateral) Conditioner This error is due to time out on Communication to the frontal (lateral) Detector Conditioner or bad status returned by Conditioner. 1200112 ERR_CONDITIONER_BOOT_WARNING Conditioner_warning [DemSupervisor::_bootConditioner] The link with the conditioner is down (Retry number: 1) followed by 1200107 ERR_CONDITIONER_BOOT_FAILURE Component(s)_Conditioner [DemSupervisor::_bootConditioner] SMC detector Conditioner ERR14 is displayed on the LCD SMC detector Conditioner ERR14 is displayed on the LCD. When viewing the status of the SMC conditioner, it displays Error code 14 on the Display. Error code 14 indicates a conditioner Alarm, the Thermostat temperatureis above 60C or less than 4C.
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Component
- Detector conditioner 5131740 SMC detector conditioner Detector conditioner Current rev is -15 but a -7 through -15 is acceptable. Detector Conditioner, connecter kit 2379172 5131740 SMC detector conditioner Detector conditioner Current rev is -15 but a -7 through -15 is acceptable. Detector Conditioner, connecter kit 2379172
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.