Pular para o conteúdo principal
GE Healthcare · MRExcite22493518275114048

TPS download problem Due to the network switch in the System Cabinet.

Symptom

Scan Stops in Middle of Scans See Long Description Hard failure during scanning, APS works ok, AP recon error watching APS window.

Error message

2249351 Exception: exception type Data Access (DSI). task number xxxxxx; task name xxxx, exception vector xxxx, pc xxxx. Host: b4 Proc: diagCoord Error:2249351 File: excLog.c Line: 670 Task ID: 0xdd9ef0 Time: 82751 Object: SCP Exception: exception type Data Access (DSI). task number 0x6E2B90; task name aceTimer, exception vector 0x300, pc 0x9E0098. Additional information in /usr/g/service/log/vxExLg-14048

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues