Pular para o conteúdo principal
GE Healthcare · MRExciteSigna224868322238172248664

TPS or MGD Reset Fails With MDS Link Errors due to SSM.

Symptom

TPS or MGD Reset Fails With MDS Link Errors

Error message

2248683 MDS Configuration Failed. MDS link errors caused the configuration to fail. Tue May 25 10:38:44 2004 2248683 MROP NSPSCP:MdsAutoConfig 338 MDS Configuration Failed.MDS link errors caused the configuration to fail. Please fix and perform a TPS Reset.

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2248664 MDS link failure. An MDS device failed to respond to SCP. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. 2248664 lx-mr3 NSP SCP:MdsAutoConfig 1731 MDS link failure. An MDS device failed to respond to SCP. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0000 Transmitted header: 0x11 SSM Transmitted opcode: 0x01->

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues