Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnovaAdvantx50125

The Image Intensifier/ Lift does not drive from the TSUI

Symptom

The Detector or II/ When attempting to drive the II/Detector lift from the TSUI (smart handle or smart box) It does not drive up or down.

Error message

error Code = 50125, error Text = POS TSUI: SMB or SMH error, debug Text += pos-error=4ACA, hw-status=00AC, recovery Class = DO, major Function /= LC Gantry Error Points to the SH Master - Joystick Control- II ElevatorControl Error Log had thirty two 50125 errors in two days Replaced the Smart Handle

Applies to

Models
Innova, Advantx
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Software
$, Innova 2000 Service Pack FMI F4, HerSvcPack_2.02 HerSvcPack_2.02 DL Applications version 11.23.26 Ghost DL-1 "F" : 17.4 Ghost DL-1 "G" : 15.5 Ghost DL-2 : 16.4 Ghost DL-4 : NA DL : 11.23.26 DL Atlas Core : 3.1O-1 RTAC Atlas Core : 3.1O-2 IDC : 2.13 Calypso : 11.13_FMI_7 Calypso Uptime : 11.25_FMI Titan : H0.57 Pos. Motorola : 5.2Y3-2 Pos. Roadrunner : 5.0R
Component
smart handle smart handle smart handle one type of handle is: 5145783-2 FRU Smart handle Zenith_Unity 212000947 09/22/2010 10:28 GMT see vascular suport central site (@vascparts) for a cross refference of handles, boxes, and tssc's 212000947 04/09/2010 18:48 GMT
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues