GE Healthcare · X-ray VascularInnovaAdvantx1050189
INNOVA: "Smart Handle Problem.
"Smart Handle Problem.
Symptom
INNOVA: "Smart Handle Problem. Call Service" is displayed on the Innova image review monitor.
Error message
1050189 MASTER SH GANTRY JOYSTICK TWIST ROCKER WR PosUserInterface No Boot SH_Master_gantry_joystick_twist_rocker 0 4 - Big Pivot/C-arc joystick satellite failure Z:/heroic/newsmarthandlelayer.c 3718 1050189 MASTER SH GANTRY JOYSTICK TWIST ROCKER WR PosUserInterface No Boot SH_Master_gantry_joystick_twist_rocker 0 4 - Big Pivot/C-arc joystick satellite failure Z:/heroic/newsmarthandlelayer.c 3718. Replaced the handle
Applies to
- Models
- Innova, Advantx
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Software
- $ Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9 FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 healthpage tag for this software is UNITY_M4_FMI_4.21 212009270 03/26/2010 13:31 GMT
- Component
- smart handle smart handle smart handle one type of handle is: 5145783-2 FRU Smart handle Zenith_Unity 212000947 09/22/2010 10:28 GMT see vascular suport central site (@vascparts) for a cross refference of handles, boxes, and tssc's 212000947 04/09/2010 18:48 GMT
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.