GE Healthcare · X-ray VascularInnova500631000014600105100025

Innova: Positioner error.

Positioner error.

Symptom

Innova: Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited, Lost All Gantry axis's drive or NO movements. Omega table elevation will not drive down. Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited, Lost All Gantry axis's drive or NO movements. Omega table elevation will not drive down. Lift failure, dose is minimized, call service

Error message

INX100: 50063 GENERAL POSITIONER MODULE SOFTWARE ERROR ( 4A70 ) Followed by 1000014, 600105 and 100025 Arcnet failures 50063 POS : GENERAL POSITIONER MODULE SOFTWARE ERROR DO VAS / POS (null) Software Problem 1 1 POS ERROR=4A70,HDW STATUS=0031,USR STATUS=0001 XRPAtlasLogger.cpp 99 Followed by 1000014 SID SSD out of range 600105 ARCNET Communication Failure

50063 POS : GENERAL POSITIONER MODULE SOFTWARE ERROR HWStatus=0032 User=0000 0001 4A70

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Software
$, Innova 2000 Service Pack FMI F4, HerSvcPack_2.02 HerSvcPack_2.02 DL Applications version 11.23.26 Ghost DL-1 "F" : 17.4 Ghost DL-1 "G" : 15.5 Ghost DL-2 : 16.4 Ghost DL-4 : NA DL : 11.23.26 DL Atlas Core : 3.1O-1 RTAC Atlas Core : 3.1O-2 IDC : 2.13 Calypso : 11.13_FMI_7 Calypso Uptime : 11.25_FMI Titan : H0.57 Pos. Motorola : 5.2Y3-2 Pos. Roadrunner : 5.0R
Component
CPU Roadunner in the Positioner CPU Roadunner in the Positioner. Check FeTools and ASM for any updates on part numbers. Aways load SW on new board. Part #: 5145126 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 2.43 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: RP CPU ROADRUNNER CVC FRU: Yes Status: Orderable CPU Roadunner in the Positioner. Check FeTools and ASM for any updates on part numbers. When replacing CPU, Aways load SW on new board. Part #: 5145126 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 2.43 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: RP CPU ROADRUNNER CVC FRU: Yes Status: Orderable
Area
X-ray Vascular