Pular para o conteúdo principal
GE Healthcare · MRExcite22238172252106

2223817 TPS is Not Responding Due to Reflex 400B Minerva Master AP Board.

Symptom

TPS not responding during applications boot with Error 2223817 See long description and attached file. NO errors in mgd_term windows. All three processors finish boot with applications startup complete message. /usr/g/service/log/scn.out file showed MGD chassis failed boot. See attached file.

Error message

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2252106 Failed to determine hardware state for APS board within 20 seconds. Error: 2252106 Host: BARLOW_MR Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for APS board within 20 seconds The board may be busy loading vxWorks or may be hung. Also check the serial cable for the board and the serial terminal server.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2325947-12 AP Reflex Minerva II 400B Master Board - M3000RJ For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required