GE Healthcare · MRSignaExcite2247383685114
Defective UTNS3 was verfied as causing error 2247383, swapped UTNS3 in slot 12 with UTNS3 slot 3 and problem moved.
Symptom
16 Channel Switch Diagnostics Fail
Error message
2247383 The signal's channel-to-channel variance exceeded the specification for channel: Test 16 Channel Switch Output Osc See Long Description Error: 2247383 Host: b5 Process: diagCoord File: rrfRxFruIso.c Line: 2796 Diagnostic: 16 Channel Switch Iteration: 1 of 1 Action: 1 Task ID: 0x7f53e50 Time: 685114 Object: APS Test Task Function: Software/Firmware/Config: N/A Test: '16 Ch Switch Output Osc. OB(O)' diagnostic error. The signal's channel-to-channel variance exceeded the specification for channel: 7, 8.-> Extended Error: Review the data in /usr/g/service/datamine. The summary file contains the results used for logging this error. If the file exists, the raw and processed data is available in a corresponding data file whose name contains the test iteration tag.
Applies to
- Models
- Signa, Excite
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2366218-2 UTNS3 See Long Description 2366218-2 UTNS3 EMC ED2 Compliant .
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.