Pular para o conteúdo principal
GE Healthcare · MRExcite222384922238592252106

Intermittent Boot up Failure and Intermittent Scan Pause.

Symptom

Intermittent Boot up Failure and Intermittent Scan Pause The MGD intermittently fails to boot. When it was successful in booting software after a TPS reset, the scan would sometimes pause without operator interaction.

Error message

2223849 AGP EPROM power-up boot failed See Long Description Error: 2223849 Host: rmns Process: NSP File: TardisReset.cc Line: 1781 AGP EPROM power-up boot failed. Communication Timed-out from AGP board. Verify that MCP820 boards are installed and seated properly for APS and AGP. Extended Error : MCP820 boards can be identified by the word MCP820 printed on the board extraction lever.

2223859 The TPS Reset Failed because of a VRE Reset Failure See long description. 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Extended Error: This indicates that a failure to reset the VRE properly caused the TPS Reset to fail. Please refer to the VRE Troubleshooting Guide for further troubleshooting this issue, which can be found in the MR Service Documentation.

2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
5139480-2 8 PORT GIGABIT ETHERNET SWITCHING HUB See long description for important note FEMC Description: 8-Port Switch Kit with 4 cushions To view the most recent appropriate and detailed documentation, click to go to SERVICE METHODS. Note: BE CAREFUL when ordering this part. Per the FRU manuals, some products use this part for GOC and/or System Cabinet and some don't. Some products are supposed to use the 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT as the FRU. Please check the FRU manual for your system to see whether this part should really be used on the Product and for the purpose (i.e. GOC and/or Sys Cab).
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.