Pular para o conteúdo principal
GE Healthcare · MRExcite2248108

Signal was Too Large to Process Due to Port A plus Bazel and Cable Harness.

Symptom

Auto Prescan fails with Signal was too large to process. Use Manual Prescan. Auto Prescan failed. Signal was too large to process. Use Manual Prescan.

Error message

2248108 APS received data over range error from DRF, on chan= n. 2248108 TOiMR2 NSP APS:dataMoverTask 604 Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 APS received data over range error from DRF, on chan = 2 Need to reduce receiver gain.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
5111020-50 HDMR2 A plus Bezel and cable harness.
Area
MR