GE Healthcare · MR

GE supplied chiller still locked due to registration process not completed.On new installs looking for the code to unloc…

GE supplied chiller still locked due to registration process not completed.On new installs looking for the code to unlock their chiller - should those go to OLE or page FE?Page the FE for questions ab…

Symptom

On new installs looking for the code to unlock their chiller - should those go to OLE or page FE? Some GE-supplied chillers have not been registered at the time of install and may still have the Lock attached. Failure to complete registration may void chiller warranty.

Applies to

Product
Magnet LCC. Software Version Concept should NOT be selected for Magnet Products Serial #: 1.5T (Rxxx) / 1.0T (Qxxx) He Vessel Pressure:>3 psig but <5 psig Shield Temp:< 55 K Coldhead Temp:> 3.8K but < 4.5K Recondenser Temp:> 4.2K but < 4.55K Delta Temp (Coldhead - Recondensor):> 0 but < 0.3K Min. LHe Fill Level: 50% Min. LHe Ramp Level: 90% Min. LHe Shim Level: 75% Helium Volume:1940 L (1.5T) Helium Volume:1970 L (1.0T) Burst Disk <":2319285 Burst Disk >":2350415 Burst Disk O-ring:46-281101P1 Baffle Assy 2180818 Low Cryogen Consumption
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.