GE Healthcare · MRExciteSigna22481021015122480142218537
FSPGR Scan Sequences Fail due to DRF1 Board.
Symptom
FSPGR Scan Fails with DPR Memory Errors The customer was unable to run FSPGR sequences. FE rebuilt protocol and tried a GE protocol but was still unable to scan FSPGR sequence.
Error message
2248102 Software communications error. Received invalid cm packet 10151, scan state = 9 2248014 Allocate DPR memory error 2248014 - Allocate DPR memory error. Possible reason : too many printf or log msg caused APS missed EOS interrupt. Software error. Please change prescription or perform a TPS reset. DRF ID : 4 FE ID : 0 arg0 : 1 arg1 : 0 2218537 Scan stopped, Got Recon-Scan Stop packet With Error Message Key Fri Apr 16 10:11:34 2004 Error: 2218537 Host: LX1 Process: scn File: MsaScanMgr.m Line: 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2248
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2215142 DRF board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.