Pular para o conteúdo principal
GE Healthcare · MRExciteSigna22481021015122480142218537

FSPGR Scan Sequences Fail due to DRF1 Board.

Symptom

FSPGR Scan Fails with DPR Memory Errors The customer was unable to run FSPGR sequences. FE rebuilt protocol and tried a GE protocol but was still unable to scan FSPGR sequence.

Error message

2248102 Software communications error. Received invalid cm packet 10151, scan state = 9

2248014 Allocate DPR memory error 2248014 - Allocate DPR memory error. Possible reason : too many printf or log msg caused APS missed EOS interrupt. Software error. Please change prescription or perform a TPS reset. DRF ID : 4 FE ID : 0 arg0 : 1 arg1 : 0

2218537 Scan stopped, Got Recon-Scan Stop packet With Error Message Key Fri Apr 16 10:11:34 2004 Error: 2218537 Host: LX1 Process: scn File: MsaScanMgr.m Line: 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2248

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2215142 DRF board
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues