GE Healthcare · X-ray R&F17025
PROTEUS FMI 17025 and new installation issue
Symptom
PROTEUS Exposure hold. There is an issue with both FMI 17025 and new Proteus installations using 2.13 version console software. There can be strange exposure hold issues with no error messages. Typical scenario: Table Bucky is selected, cassette in tray, and OTS at the proper SID. The system will produce an exposure hold. If the customer chooses table top and then reselects the table bucky, exposures are permitted until such time as the cassette is removed and reinserted.
Applies to
- Product
- PROTEUS XRa
- Software
- Proteus Console Software version 2.13
- Component
- Table Bucky Wall Bucky Proteus Console
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.