GE Healthcare · MRExcite2254793
UPM Fails DAC Self Test Due to UPM detector board.
Symptom
UPM Errors During Scan
Error message
2254793 The Universal Power Monitor 1 failed its DAC Self Test on the Narrow Band Head Channel. See Long Description 2254793 RDC NSP SCP:UPM 1059 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: Unknown Coil: Unknown Scan: -1/-1 The Universal Power Monitor 1 failed its DAC Self Test on the Narrow Band Head Channel. Expected Data : 0x3c00. Detected Data : 0x7fff.->
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5250034 RF Detector 2 Board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.