UPM Functional Checks do not run properly.

Symptom

UPM Functional Checks do not run properly See Long Description Calibration Wizard shows UPM functional checks ( no cal's). When we run functional check, it fails with indications the system is not cal'ed. When running Functional Check, we get errors in the in test's dialog box saying Peak power too low; pulse width too low, etc., but test continues to run. The test will stop with a message in the dialog box to check the error log for a "pulse width trip" message, however we do not get that error. Instead we get RF error 2254055 Rrecovery in Progress and error 2256853 ACE Timeout error ( maybe we did not answer the dialog box in a certain amount of time?).

Error message

2256853 The ACE Timer monitor timed out. 2256853 lx-mr LxMxConsumer MonitorHost.cpp 366 Server Name: SC The ACE Timer monitor timed out.

2254055 Recovery in progress, see log

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Area
MR

Full cause & step-by-step resolution