GE Healthcare · MRExcite22180622224382
Illegal Character In The Localizer Protocol Name.
Symptom
MRIQ Black Images No Signal See image Black images, no signal.
Error message
2218062 Unable to add image path/header to tuple list Thu Oct 9 11:00:07 2003 Error: 2218062 Host: lx-mr Process: TIR File: ImageInstall.cc Line: 416 Unable to add image path/header to tuple list 2224382 DICOM Translation Failed Thu Oct 9 11:00:07 2003 Error: 2224382 Host: lx-mr Process: TIR File: storeImage.c Line: 537 DICOM Translation Failed for Ex 385 Se 1 Im 1, Element (0x18, 0x1030). Note: Dicom translation error points to a problem with the (0018 1030) LO Protocol Name.
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution