GE Healthcare · MRExcite22180622224382

Illegal Character In The Localizer Protocol Name.

Symptom

MRIQ Black Images No Signal See image Black images, no signal.

Error message

2218062 Unable to add image path/header to tuple list Thu Oct 9 11:00:07 2003 Error: 2218062 Host: lx-mr Process: TIR File: ImageInstall.cc Line: 416 Unable to add image path/header to tuple list

2224382 DICOM Translation Failed Thu Oct 9 11:00:07 2003 Error: 2224382 Host: lx-mr Process: TIR File: storeImage.c Line: 537 DICOM Translation Failed for Ex 385 Se 1 Im 1, Element (0x18, 0x1030). Note: Dicom translation error points to a problem with the (0018 1030) LO Protocol Name.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR