GE Healthcare · MRExcite2252246
Will Not Scan with Gradient Errors due to grafidy cal files not present.
Symptom
Will Not Scan with Gradient Errors.
Error message
2252246 Quick reset detected on the CAN network. Recovered Node is Master Gradient 2252246 NSP SCP:CanLinkHWO 799 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 Quick reset detected on the CAN network. The node reset and recovered quickly. Recovered Node is Master Gradient.-> See Long Description for extended version of error message. 2252246 NSP SCP:CanLinkHWO 799 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 Quick reset detected on the CAN network. The node reset and recovered quickly. Recovered Node is Master Gradient.-> Extended Error: The CAN node reset and recovered within a short amount of time. This message is expected if a CAN node was manually reset. If not manually reset, check that the CAN and power cable connections to the node are solid, and that the power supply to the node is stable.
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.