GE Healthcare · X-ray VascularInnova60008
No gantry motion is reported by the customer.
Symptom
No gantry motion is reported by the customer. The Gantry does not drive. Emergency backout works
Error message
60008 POS : GANTRY INIT FAULT. CALL SERVICE. HWStatus=000x User=0000 000x 4AE1 60008 POS : GANTRY INIT FAULT. CALL SERVICE. WR GANTRY 0 0 HWStatus=000x User=0000 000x 4AE1 HW status: 000x where x = 0000 L–arm smart–amp board 0001 LC pivot smart–amp board 0002 LC arc smart–amp board 0003 LC II smart–amp board 0004 LP II smart–amp board 0005 LP X–ray Tube smart–amp board 0006 LP pivot smart–amp board 0007 LP arc smart–amp board 0008 System interface board 0009 Omega–4 stepping Ctrl 60008 POS : GANTRY INIT FAULT. CALL SERVICE. WR GANTRY 0 0 HWStatus=000x User=0000 000x 4AE1 HW status: 000x where x = 0000 L–arm smart–amp board 0001 LC pivot smart–amp board 0002 LC arc smart–amp board 0003 LC II smart–amp board 0004 LP II smart–amp board 0005 LP X–ray Tube smart–amp board 0006 LP pivot smart–amp board 0007 LP arc smart–amp board 0008 System interface board 0009 Omega–4 stepping Ctrl Get the user code and procede to troubleshooting. User codes: 0001: PRD failure. 0002: Too many communication errors during board setup. 0004: Data setup verify error. 0008: Board deactivated by its associated “buddy board”. 0081: 80C196 does not respond to initialization or instructions not correctly execut
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- Positioner Gantry Positioner Gantry Positioner Gantry
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.