Pular para o conteúdo principal
GE Healthcare · MRSignaExcite222570422259352225334222569122253372225697

MDS Link Failure on First Bootup of the Day due to CPD Module.

Symptom

MDS Link Failure on First Bootup of the Day See Long Description MDS link failures on 1st boot up of the day, every day. Power is shut off at A1 box every night. Customer simply reboots to recover. The problem only happens when system sits overnight with no power. The InHouse II engineer can't duplicate the problem during the day, even with A1 power off. The system also logs PAC and SRI not responding errors, which implies no power to them from the SSM. Unknown if power is being applied to them or not. Error codes logged are 2225691, 2225697, 2225334, 2225337, 2225704, 2225935 , with " MDS Link is broke between the RF and Syscab ". (see Error concepts for specifics on these errors). In the MDS link failure errors, this code is always seen, which decodes as the RF cabinet, on this system that would be the CPD Module in the SSM: Transmitted byte: 0x11 Author Pual Lyman SME Scott Allsop October 12, 2009

Error message

2225704 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Error: 2225704 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_mds Line: 300 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.

2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. 2225935 The PAC does not respond. Check PAC power and fiber optic connections, run PAC diags. This error condition adversely affects gated scans only. 2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. This error condition adversely affects gated scans only.

2225334 SPI-SRI Communication Error. SRI failed to respond to a data packet from SPI. Error: 2225334 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri resp timeout Line: 12 SPI-SRI Communication Error. SRI failed to respond to a data packet from SPI. Data packets waiting for a response: 0x4000

2225691 MDS link failure. The IPG failed to receive an MDS transmission

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion CPD Module 2160200-5 This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.