-06- and -03- Displayed on the Magnet Display.
Symptom
-06- and -03- Displayed on the Magnet Display See Long Description -06- on magnet display. Cradle only moves in the out direction because it doesn't see home even though the Table Interface indicates it's at home (home switch is good). The Relay Board sees all switch action from the Magnet Control panel which indicates cabling/communication is good from the magnet to Relay Board is good. -06- error timed out then immediately replaced by -03- error on the Magnet Display. Depending on cradle position, either -06- or -03- is displayed. Both yellow cradle unlatch switches are in which engages cradle drive. If one or both switches are out, the cradle is unlatched from the drive motor.
Error message
-06- Error Displayed on the Magnet Display. Cradle accuracy slippage This error is not recorded in the system log, only displayed on the Magnet Display. -03- Error Displayed on the Magnet Display. Cradle is not Latched. This error is not recorded in the system log, only displayed on the Magnet Display.
Applies to
- Product
- Profile 3 Ultrasparc CPU with AN8290Y gradients
- Software
- Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
- Component
- 2161974 or 5114871-2 SRP Assy 2161974 upgraded to 5114871-2. Same functionality as an SRI. See SN DOC0625410 Rev.1 dated July 7, 2009.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.