Propeller Scan will Not Run due to Defective Reflex MINERVA AP (SLAVE).

Symptom

Propeller Scan Option Will Not Run or Runs Very Slow propeller option will not function

Error message

2257001 This application cannot be run on this system due to insufficient image reconstruction power.

2253517 The system did not find the right amount of AP memory. Only 1212 Meg was found when there should be 1624 Meg. This indicates that some AP board memory is not working.-> See Long Description 2253517 2 WFBIMRI diagCoord apBld.c 315 Diagnostic: Dual Board AP Mem Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6948 Time: 16436 Object: APS Test Task The system did not find the right amount of AP memory. Only 1212 Meg was found when there should be 1624 Meg. This indicates that some AP board memory is not working.->

2253518 The system did not find the right number of CPUs on the AP boards. Only 6 CPUs were found when there should be 8. This indicates that one of the AP boards may be partially defective.-> See Long Description 2253518 2 WFBIMRI diagCoord apBld.c 324 Diagnostic: Dual Board AP Mem Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6948 Time: 16436 Object: APS Test Task The system did not find the right number of CPUs on the AP boards. Only 6 CPUs were found when there should be 8. This indicates that one of the AP boards may be partially defective.-> Extended Error : There shoul

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
SecondOpinion MERCURY 910-07071 MINERVA II SLAVE 2325947-13 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.