Coil Not Recognized Due To Bent Pins In Quick Disconnect Box.
Symptom
Coil Not Recognized and CoilID Light is Green The coil will not scan, coil not recognized, Body Transmit Enable mismatch detected error.
Error message
2258107 Host coil validation error Body Transmit Enable mismatch detected Error: 2258107 Host: MR3T Process: coilCK File: coilCK scp.c Line: 1026 Exception Class: Unknown Function: Host: Coil Check Host coil validation error: Body Transmit Enable mismatch detected. Connector: Multicoil A Port Coil: HD Shoulder Array by GE - A/B port
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 2416877 CABLE ASM for 3T 3 CHANNEL SHOULDER COIL M3335LR 2416877 CABLE ASM for 3T 3 CHANNEL SHOULDER COIL M3335LR
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.