Fails TPS Reset when booting up to Applications or when Resetting TPS.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2247159 Sine Wave Table on RIF DIF1 Board Did Not Load Correctly See Long Description Fri Jan 25 09:23:14 2008 Error: 2247159 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 412 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51174 Object: APS Test Task The Sine Wave Table on the DIF board in RRF chassis (front, slot 10) did not load correctly. This will prevent proper Exciter output. Try TPS_reset. Run RRF diagnostics. Reported from Irf::reset_rrf (0) (0) (150)->

2248330 SCP was unable to open or read the file TnsConfig.cfg. 2248330 NSP SCP:scp_rd_cfg_file 519 SCP was unable to open or read the file TnsConfig.cfg. Verify that the file is located in /w/config. If the file is present, the /etc/vxhosts file may be missing or corrupted.

2261007 The file MRconfig.cfg contains a missing or bad value for RfBoardDriverCreator 2261007 diagCoord Irf.cpp 390 Task ID: 0x3f1fdf8 Time: 1053 Object: APS The file MRconfig.cfg contains a missing or bad value for RfBoardDriverCreator::getInstance. value = rfBoardType

Applies to

Models
Excite
Product
Profile with Excite
Software
Profile5 Applies to Profile5 (Excite) only. Profile5 is 12.x based software, not ProfileHD.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.