Pular para o conteúdo principal
GE Healthcare · MRExciteDiscovery2218527

Images Reconstructed but not Listed in the Browser, Only Displayed in Auto viewer.

Corrupt Image Database.

Symptom

Images Reconstructed but not Listed in the Browser, Only Displayed in Auto viewer. See Long Desc. Caller had images in Auto Viewer but not in Browser. Caller had scanned several series since but had turned off Autoview prior to scanning the additional new series. When Autoview is turned off, it will hold onto the last images that were scanned with it turned on.

Error message

2218527 Scanning has stopped. Image file creation unsuccessful. Please try again. Error: 2218527 Host: UPCH-MR Process: scn File: MsaScanOpsScrn.m Line: 733 Scanning has stopped. Image file creation unsuccessful. Please try again.

1036 Command Failed : TerraDBReader::ExecuteCountImages Unable To Get Image Count!!

2052 Error during execution of the fonction /export/home1/sdc_image_pool/reserved Input/output error Error: 2052 Host: lx-mr Process: imserver File: imsprim.c Line: 603 error during execution of the fonction /export/home1/sdc_image_pool/reserved Input/output error

Applies to

Models
Excite, Discovery
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR