Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova4200067

Protocol or Acquisition mode selection failed is displayed on the Innova Monitor.

Symptom

INx100: Protocol or Acquisition mode selection failed is displayed on the Innova Monitor. When trying to select another protocol or select another acquisition mode in a selected protocol, the message “Selection failed, Retry protocol selection” is displayed on live monitor. Then, X-rays are not available. Protocol or Acquisition mode selection failed is displayed on the Innova Monitor. When trying to select another protocol or select another acquisition mode in a selected protocol, the message “Selection failed, Retry protocol selection” is displayed on live monitor. X-rays are not available.

Error message

INx100: 4200067 IH XRUserIf User request failed Debug: [XRUserIf::setActiveOpModesDone] setActive opmode received with bad status.

Applies to

Models
Innova
Product
X Innova 4100 X Do not use---------4100 systems 4100 systems
Software
$ Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9 FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 healthpage tag for this software is UNITY_M4_FMI_4.21 212009270 03/26/2010 13:31 GMT
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues