Tube Spit or kV instability, steady state error Related

Symptom

JEDI: Tube Spit or kV instability, steady state error Related JEDI: Tube Spit or kV instability, steady state error Related. X-rays acquisition can be aborted if too many tube spits are occuring. JEDI: Spit or kV instability, steady state error Related. X-rays acquisition can be aborted if too many tube spits are occuring.

Error message

3602085 Jedi: too many tube spits Data Related to Error: X (Anode + Cathode Spits), AR 3602085 Jedi: too many tube spits AR XRTechMgr acquisition Jedi too many tube spits 10 JEDI Error Code: 0316H Spit Max error Data Related to Error: X (Anode + Cathode Spits) Data related to Error codes: 4,5,6,10. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service". 3602085 Jedi: too many tube spits AR XRTechMgr acquisition Jedi too many tube spits 10 JEDI Error Code: 0316H Spit Max error Data Related to Error: X (Anode + Cathode Spits). Data related to Error codes: 4,5,6,10 Review the Data related to the error. If Anode + Cathode Spits Data related to error = 4 select this concept. Data Related to error indicates the spit type: Error:1 = Anode - Ground Error: 2 = Cathode - Ground Error: 4 = Anode - Cathode Error: 8 = kV Regulation Error 10 = 8 and 2

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Ulysses MX160 X-Ray Tube Ulysses MX160 X-Ray Tube Part number 2216500 (new) or D2802A (reload) Ulysses MX160 X-Ray Tube Part number 2216500 (new) or D2802A (reload)
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.