GE Healthcare · X-ray VascularInnova3602034

Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.

Symptom

Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76

Error message

3602034 IH XRTechMgr Jedi: Heater 1 detects open circuit Debug: jedi-error-code=1211h JEDI Error Code: 1211H Heater 1: PWM rate too high Data Related to Error: 29 Tube 1 70C: OK, Tube 40C: Set, Rotor Thermal Error: OK, Inverter thermal Error OK System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service". 3602034 IH XRTechMgr Jedi: Heater 1 detects open circuit Debug: jedi-error-code=1211h JEDI Error Code: 1211H Heater 1: PWM rate too high Data Related to Error: 29 Tube 1 70C: OK, Tube 40C: Set, Rotor Thermal Error: OK, Inverter thermal Error OK System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service".

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Software
$ Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9 FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 healthpage tag for this software is UNITY_M4_FMI_4.21 212009270 03/26/2010 13:31 GMT
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues