Unable to Perform Grafidy3 with New Grafidy3 6-Coil Kits Received March and April 2009 due to 1.5T Grafidy3 6-Coil Kit 5…

Unable to Perform Grafidy3 with New Grafidy3 6-Coil Kits Received March and April 2009 due to 1.5T Grafidy3 6-Coil Kit 5135527 has Incorrectly Programmed Coilid Code.

Symptom

Unable to Perform Grafidy3 with New Grafidy3 6-Coil Kits Received March and April 2009. See Long Description Unable to perform Grafidy3 with new Grafidy3 6-Coil kits (officially known as "6-Coil ECMT Kit, 1.5T") received March and April 2009. When attaching the Grafidy3 6-Coil kit, it results in a pop-up asking for installation of coil configuration CD. No error is logged.

Error message

Popup Message "Install Coil Configuration CD" See Long Description A pop-up message appears to the operator asking for installation of coil configuration CD. No error is logged.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
5135527 1.5T Grafidy3 6-Coil Kit (6-Coil ECMT Kit, 1.5T) Officially known as "6-Coil ECMT Kit, 1.5T"
Area
MR

Full cause & step-by-step resolution